Wednesday, 21 March 2012

Muscle Uptake on MDP Bone Scan



This is an interesting scan, because not often do we come across muscle uptake in our bone scans.  For the most part we shouldn't see muscle uptake at all, however it is not uncommon, and there has been several journaled articles relating to this fact.  Having "googled" the topic, it appears that there are several conditions that can cause this extra osseous uptake.  There has also even been one case in which exercise prior to the bone scan has caused muscle uptake.

On a quick note, here are some interesting articles (Radionuclide Bone Imaging:  An Illustrative Review) in regards to screening pathological conditions and as well Nonosseous, Nonurologic Uptake on Bone Scintigraphy:  Atlas and Analysis that provides some of the pathways in which Tc99m MDP are involved with non osseous uptake.

With this patient here, you can see if you expand the image, that there are diffuse muscle and soft tissue uptake throughout the body:  hip flexors, quadriceps, calf muscles, deltoids, triceps  and breast tissue.  The patient had been suffering from left sided flank pain and had undergone several examinations to determine the nature of the pain.  This person also has had a previous liver transplant and heart transplant as the result of being diagnosed with glycogen storage disease (GSD) type IIIA at an early age.  We are not really sure how the Tc99m MDP got into the muscles and other soft tissues, but conditions such as polymyositis, myositis ossificans or amyloidosis can cause this appearance.  However there are a slew of conditions that can also cause this, some of which may be pathological in nature whereas other may be more technical, like too much reduced hydrolyzed in the radiopharmaceutical.

In the end, the left sided flank pain may have been attributed to a 12th rib fracture, as noted on the bone scan.  A SPECT acquisition was also performed to determine the location.  Furthermore in the final report, the diffuse uptake in soft tissue was discussed with the referring physician about the possibility of an underlying myositis or anasarca.



Monday, 19 March 2012

Something Doesn't Look Right!


Fig. 1  This is an anterior TBBS image, but something doesn't seem quite right.


Fig.2  The posterior TBBS image.  Expand the image.


When we first looked at these images we were unsure what the problem was, but the images did not look right.  The total body bones scan was taken on a Philips ADAC Skylight system, but generally the images do not look this way.  This is what I mean.... the images look hazy and not indicative of a normal quality bone scan. Some of you may have figured it out by now, but for those who are still guessing, it was caused by a student.

The patient had come to the Nuclear Medicine department to determine if osteomyelitis had developed in the right shoulder.  They have had several corrective surgeries to their joints already:  bilateral shoulder replacement, bilateral hip replacement and a left knee replacement.  Let's put it this way, when she came into the department the patient was held together by duct tape... joking of course but the patient was in pain.

However knowing her condition, we had to repeat the images because something was amiss.  Check out the images below, which were also performed on the Philips ADAC system.



Fig. 3  Corrected TBBS images.  Compare to Fig. 1 and Fig. 2.  Expand the image.

So the problem is this... when the student had placed the patient on the bed, they had unknowingly changed the collimators to high energy collimators instead of the high resolution collimators (VXGP).  While setting up the patient in the computer for the acquisition, the computer had prompted the student that the wrong collimators were installed.  Majority of our acquisition protocols are customized to the organ system that we are scanning (like most modern departments), with a predetermined energy spectrum and energy window, collimation, length of time for the scan etc.. In this case the student did not bother to read the prompt which notified that the wrong collimators were installed for this particular acquisition protocol.  They ignored it and proceeded with the scan.  Thus the 140 keV gamma rays hitting the crystal were limited by the thicker collimation of the high energy collimators, since they were designed for high energy isotopes to minimize the amount of cross talk.  Collimation is one of the fundamental principles of Nuclear Medicine Instrumentation.

Fig. 4  An example of a parallel hole collimator.  With high energy collimators the "holes" are smaller due to the thicker septa and the longer bore, thus limiting the amount of detection or the sensitivity of the collimator.  The reason for this is to limit the amount of cross talk when imaging with high energy isotopes like iodine.  If these modifications were not present, it would increase the amount of scatter and thus decrease the resolution of the image.

Another possible way of getting images to look like those in Fig. 1 and Fig. 2 is by having the wrong energy or energy window settings.  Although a lot of the camera systems all now have presets, but it's always a good habit to check your parameters and never ignore your error messages.

In the end this patient had their total body bone scan, along with their In-111 WBC with sulphur colloid two days later and found no osteomyelitis in the right shoulder.