Friday, 9 March 2012

I-131 Uptake Post Therapy

Fig 1.  Radioactive iodine uptake midline to the body.  The scan was performed 10 days post iodine therapy.

A patient was presented to our department recently with papillary carcinoma.  The right lobe was removed in the early 1990's while the left lobe was removed in 2011. Follow up treatment was provided by administering 7500 MBq of I-131. Ten days after the therapy, a whole body iodine scan was presented with a focal uptake, midline to the body.  By the nature of the location, it's a bit tricky because we want to know if this is focal bowel activity (diverticulum ?) or is it a true lesion (met?).

There are a couple things that we can do to differentiate, or at least provide some clues,  in determining whether the "spot" is  bowel or something else.  They are the following:

1.  SPECT the area of concern to determine where it is within the body and try to correlate this information with any prior or current imaging "work ups".

2.   Acquire laterals, if the SPECT resolution is poor, and compare with correlative imaging.

3.  Have the patient return the department the next day and repeat the image over the midline to see if the "spot" would move.  This would help us to determine if this was bowel activity.

4.  Perform a SPECT/CT of the area to localize the "spot".

Lucky for us we do have a SPECT/CT in the department, and this is what we have acquired.

Fig. 2  Fused coronal image.

Fig. 3 Fused sagittal image.

Fig. 4  Fused transaxial image.

These are the fused images, since stand alone SPECT images of the site does not really provide as much information in comparison to SPECT/CT, because we can't locate where the "spot" is in relation to hard physical anatomy.

Regular biodistribution of iodine includes salivary, nasal-oral, hepatic, bowel and mammary uptake.  Having said that, the radioactive uptake seemed to be quite focal on both the anterior and posterior images, so it was suspicious from the beginning.  Furthermore, if you look closely at figure 1, a "star artifact" (expand the image) can be seen as well.  All in all, the "spot" was suspicious and it revealed a localized metastasis on the lumbar spine. A compression fracture which was also seen on the CT of the area may have contributed to the uptake as well (ie. inflammatory responses) but the patient did not complain of any discomfort.  

Furthermore a follow up biopsy of the spine (L3), confirmed bony involvement.

Friday, 2 March 2012

How Does Attenuation Correction Work?

I think it's one of those questions where you know the answer, but really don't know the answer, because you know the reason why we do attenuation correction (AC), but really don't know how it is applied on our patients.  Thus the question, "how does attenuation correction work?"  

To start, the principles of AC applies to both SPECT/CT and PET/CT alike, even though we are using different isotopes and energy levels for these systems.  

The CT unit is the main equipment piece that helps us to make this work.  The reason why is, it is the x-ray's that shoot through the body, which gets collected by the CT detector array on the other side that provides the AC data.... "but we know that already!".  The problem lies in the fact that the x-rays and the energy of the isotopes are different.  Let's use Tc-99m as an example.  The problem is this, how do you use a 70 keV x-ray to correct for the nuclear medicine SPECT scan if the energy level for acquisition is at 140 keV?  These energies interact and attenuate differently as it traverses through the body!

Fig. 1  At the very top is the CT x-ray tube (rectangle), with x-rays shooting outwards in a fan beam array through the body (ellipse), being received by the detectors arrange in an "U shape" at the bottom.


When the x-rays are produced from the x-ray tube, there are a spectrum of low to high energies generally, where low energy x-rays get absorbed by the tissues whereas the high energy x-rays pass right through the body, if it is used unfiltered. This is something that we do not want because it makes it harder to figure out the AC maps. Thus the x-rays are filtered (wedge or bow tie) to produce an average of 70 keV's to help simplify the process.  The process isn't perfect, but it "hardens" the beam to eliminate the "ends" of the x-ray spectrum.

Fig. 2  The the x-rays are filtered to help simplify the process of AC.

Once we collect the x-ray data from the detectors, the information is reprocessed. The most important piece of information is the linear attenuation coefficient (u), which is calculated by using this formula, aka Beer's Law:

Formula taken from CT1 course, CAMRT

The little (u) can now be used to calculate the CT numbers to differentiate the different densities in the body based on the Hounsfield unit scale on our display monitors.



So the question now is, "How do we put all this information together to understand AC?!?!"

We've acquired our CT and what you see are whole bunch of CT numbers converted to the Hounsfield's scale, based on the (u) values of the CT x-rays.  The next step is to convert this data and create a correction factor for the corresponding voxels in the SPECT scan.  To do this, we need to calculate the new (u) value that is scaled to 140 keV.  Basically, what would be the new (u) value if we converted 70 keV x-ray to 140 keV (the Tc-99m energy that we want to scale it to)?

There's two parts to this scaling:

1.  Formulas to calculate and scale the 70 keV to 140 keV based on the original CT number acquired for your CT scan

2.  A bilinear graph (model) used to convert CT numbers to (u) values for specific radionuclides like Tc-99m, In-111, Ga-67, F-18 etc.

The formulas:

Fig.3  Top:  Formula for (u) for CT values less than 0 (0 being water).  Bottom:  Formula for (u) for CT values greater than 0.  These are bit hard to see, but look them up in this article:  SPECT/CT Physical Principles and Attenuation Correction

The bilinear graph, works hand in hand with the formulas, because it just gives a graphical overview of how the CT numbers are treated when the CT values are below or above CT=0.  The reason for the bilinear graph is to account for the different densities in the body (ie. water and air mix OR water, soft tissue and bone mix).

Fig. 4  Bilinear graph depicting the corresponding (u) values in relation to the original CT numbers based on the CT scan.


Fortunately most of these scaling values have been pre-calculated and installed in the computer in a stored look up table for the reconstruction algorithm... it's been all worked out. To finish it all off, the CT image has to be "dumbed down" to fit the overlying SPECT image, since CT images have higher resolution than nuclear medicine images thus it needs to be reformatted so that it matches the same matrix, slice width and slice position as the SPECT data.  The new scaled AC map is then incorporated into the original radionuclide SPECT projection image, and reconstructed.  The AC map is used to correct the emission counts from the uncorrected SPECT data to provide the final corrected SPECT data.